รายละเอียดเอกสาร

เคมี

บทความ

สหราชอาณาจักร

ไม่ระบุ

Efficacy Comparison of an Automated Growth-Based System and Plate-Count Method on the Detection of Yeasts and Molds in Personal Care Products

2023

10

Journal of AOAC INTERNATIONAL Vol. 106, No. 6, November-December 2023, p. 1564-1573

Objective
The present investigation studies the efficacy of an automated growth-based system for a quantitative determination of Candida albicans and Aspergillus brasiliensis in several personal care products. The main purpose of this validation study was to prove that the alternative method’s entire performance is not inferior to the conventional pour-plate method for a quantitative determination of yeasts and molds. Thus, a performance equivalence was established in accordance with the United Stated Pharmacopeia (USP-NF) Validation of Alternative Microbiological Methods ˂1223˃.
Methods
C. albicans and A. brasiliensis were pooled to use as inoculum (equivalent to 1.0 × 108 CFU/mL) in the suitability of the method test. PCP's preservatives were chemically neutralized leading to the yeast and mold recovery by means of the alternative microbiological method (AMM) and the pour-plate method. A correlation curve was generated for each PCP by plotting DTs relative to the corresponding log CFU values.
Results
Thirty PCPs have been tested for quantification of yeasts and molds using an AMM. An equivalence of results was made through the construction of correlation curves that allowed the establishment of numerically equivalent results between the enumeration data from the reference method (CFU) and the alternative method (Detection times, DTs). Thus, following the guidelines of USP Ch.1223, essential validation parameters were tested, such as equivalence of results (Correlation coeficient, CC >0.95), linearity (R2 >0.9025), accuracy (% recovery >70%), operating range, precision (CV <35%), ruggedness (one-way ANOVA, P > 0.05), specificity, LOD, and LOQ.
Conclusion
It was shown that all the test results obtained from the alternative method were in statistical agreement with the standard plate-count method (PCM). Thus, this new technology was found to meet all the validation criteria needed to be considered for an alternative method for yeast and mold quantification in the PCPs tested.
Highlights
In accordance with the United Stated Pharmacopeia (USP-NF) Validation of Alternative Microbiological Methods ˂1223˃, the implementation of alternative methods can offer benefits in execution and automation while improving accuracy, sensitivity, and precision and reduce the microbiological process time compared to the traditional ones.